Technical data
-80 ... + 200 °C
Wet bulb temperature
Absolute humidity
Mixing ratio
Enthalpy
-10 ... +50 °C
0 ... 20g/m³
0 ... 100g/kg dry air
0 ... 80kJ/kg
0...10 V
4...20 mA
RS232, RS485 Modbus
15 ... 30 V DC / 13...26 VAC
10 ... 30 V DC
5 ... 30 V DC
Downloads
Accessories
AW with remote probe
Transmitters for humidity and temperature can be combined with different probes
Features
- Operating temp. up to 200 °C
- Accuracy : ±1,5 % RH
- On-site calibration
- IP65
- Stainless steel sensor tube
- Modulair, can be combined
Options
The transmitter AW, (wall mounted), can be combined with any probe of the S series. The AW-transmitter combined with the probe SZKA.0H (pictured here) can measure relative humidity and temperature in operating temperatures up to 200 °C. Combined with the probe SZKA.HD in operating temperatures up to 160 °C and at pressures of up to 25 bar. Additionally there is the option for resistance to ammonia.
The integrated hx processor can calculate further humidity values and apply them to the two outputs. With the RS485 standard all of the hx-values can be read simultaneously. Due to the digitisation of signal processing, the measurement accuracy for humidity achieves excellent values of ±1.5 % RH. In temperature measurement, the platinum resistance sensor achieves tolerances of ±0.15 K.
Capacitive MELA® sensorchips
Equipped with the highly dynamic capacitive MELA® sensor element, the sensor provides reliable performance across the full humidity range. Using the MELA® sensor element ensures long-term stability, resistance to dew formation, minimal hysteresis, and strong dynamic capabilities.