Technical data

Relative Humidity
0...100 % RH
Temperature
-60 ... + 160 °C
-80 ... + 200 °C
Dew point temperature
Wet bulb temperature
Absolute humidity
Mixing ratio
Enthalpy
-20 ... +70 °C
-10 ... +50 °C
0 ... 20g/m³
0 ... 100g/kg dry air
0 ... 80kJ/kg
Signal output
0...1 V
0...10 V
4...20 mA
RS232, RS485 Modbus
Supply voltage
6 ... 30 V DC / 6...26 VAC
15 ... 30 V DC / 13...26 VAC
10 ... 30 V DC
5 ... 30 V DC

AW with remote probe

Transmitters for humidity and temperature can be combined with different probes

Features

  • Operating temp. up to 200 °C
  • Accuracy : ±1,5 % RH
  • On-site calibration
  • IP65
  • Stainless steel sensor tube
  • Modulair, can be combined

Options

  • Modbus
  • Display
  • hx converter
  • Pressure-resistant up to 25 bar
  • Pressure-resistant up to 10 bar
  • Ammonia-resistant
  • Exchangeable

The transmitter AW, (wall mounted), can be combined with any probe of the S series. The AW-transmitter combined with the probe SZKA.0H (pictured here) can measure relative humidity and temperature in operating temperatures up to 200 °C. Combined with the probe SZKA.HD in operating temperatures up to 160 °C and at pressures of up to 25 bar. Additionally there is the option for resistance to ammonia.

The integrated hx processor can calculate further humidity values and apply them to the two outputs. With the RS485 standard all of the hx-values can be read simultaneously. Due to the digitisation of signal processing, the measurement accuracy for humidity achieves excellent values of ±1.5 % RH. In temperature measurement, the platinum resistance sensor achieves tolerances of ±0.15 K.

Capacitive MELA® sensorchips
Equipped with the highly dynamic capacitive MELA® sensor element, the sensor provides reliable performance across the full humidity range. Using the MELA® sensor element ensures long-term stability, resistance to dew formation, minimal hysteresis, and strong dynamic capabilities.

Often used in

  • Agriculture & food industry
  • Chemical industry
  • Clean rooms
  • Climate chambers
  • Drying plants
  • Paint shops
  • Paper & print
  • Pharmaceutical industry
  • Process and factory automation
Cookies
Diese Website verwendet funktionale und analytische Cookies. Funktionale Cookies sind für die
einwandfreie Funktion der Website erforderlich. Sie können Ihre Präferenzen jederzeit ändern.
Informationen zur Datenverarbeitung finden Sie in unserer Datenschutzerklärung
Cookie-Einstellungen ändern
Funktionale: Funktionale Cookies sind für die einwandfreie Funktion der Website erforderlich.
Analytische: Diese Technologien ermöglichen es uns, die Nutzung der Website zu analysieren, um die Leistung zu messen und zu verbessern.
Marketing: Diese Technologien werden von Werbetreibenden verwendet, um Anzeigen zu schalten, die für Ihre Interessen relevant sind.